June 4, 2020
>> Infusion Pumps and Suction >> Infusion Pump Units

Baxter Flo-Gard 6201 Infusion Pump

Condition: Refurbished

Part #: 6201
Warranty: 6 months
(Included accessories' warranties may be limited.)

$990.00 $615.00

+ Shipping (Enter Zip Code)

Quantity:
Add to cart

Description

The Baxter Flo-Gard 6201 Infusion Pump single infusion pump is an all-purpose IV pumps that use standard IV administration sets from Baxter. The pumps feature automatic piggybacking, upstream and downstream occlusion detectors, air-in-line detection, automatic restart after occlusion clears and IV set free-flow protection.

  • The Baxter Flo-Gard 6201 Infusion Pump is a Linear Peristaltic Volumetric Infusion Pump
  • Keep Vein Open (KVO) Rate: 5 mL/hr or programmed rate, whichever is less
  • Nurse Call: Available through the COMMUNICATIONS PORT
  • Battery: 12 Volt, 2.0 Ah sealed lead-acid
  • Battery Life: Approximately 6 hours with pump running at a rate from 1 to 1400 mL/hr
  • Battery Recharge: 8 hours for more than 80% recharge
  • AC Power Requirements: 110/120V, 60 Hz
  • Power Cord: 2.9 m (9′) long, with hospital grade plug
  • Fuse: 0.5 A, 250V, SB, 5.2 mm (13/64″) x 20 mm (25/32″)
  • Leakage Current: Less than 50 microamps (using UL-544 specified test methods)
  • Weight: Approximately 5.3 kg (11.6 lbs)
  • Dimensions: 20 cm W x 13 cm D x 29 cm H (7.9″ W x 5.1″ D x 11.4″ H)

Occlusion Prevention

  • Air-in-Line Detection: Factory set to NORM, which causes the pump to alarm on air bubbles approximately 75 uL or larger. The MIN setting causes the pump to alarm on air bubbles approximately 50 ul or larger.

Delivery Rate Range

  • Flow Rate Range:
    • Primary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-1999 mL/hr in 1 mL increments. Upper limit can be reduced by authorized service personnel.
    • Secondary program: 1.0-99.9 mL/hr in 0.1 mL increments or 1-999 mL/hr in 1 mL increments
  • Ramp PDP Parameter Ranges:
    • Total VTBI: 100-9,999 in 1 mL increments. Upper limit can be reduced by authorized service personnel
    • Delay Time: 00:00-08:00 hours in minute increments
    • Run Time: 01:00-16:00 hours in minute increments
    • Ramp Up: 00:00 to 45% of the Run Time, in minute increments
    • Ramp Down: 00-00 to the Run Time minus Ramp Up, in minute Increments

Dose Limit Range

  • VTBI Range: 1.0-99.9 mL in 0.1 mL increments or 1-9999 mL in 1 mL increments for both primary and secondary. Upper limit can be reduced by authorized service personnel.



Savings of refurbished items are based on competitors' prices. Savings of new items are based on MSRP. For expedited shipping or shipping discounts for large quantity orders please call or email for quote. International Orders - For countries other than the United States, please call or email to set up your shipping rates. Customer's Shipping Accounts - If you would like us to use your shipping account please contact us. Orders usually ship within 2-7 business days. Should there be any delays in your order you will be notified with an updated shipping period. Due to continuing product innovations, specifications and packaging are subject to change without notice. Images are representative of the item and may contain options not included. Trademarks are property of their respective owners.